EE-100A-3S Three-Phase Current Load with Energy Standard Class 0.1

Three-Phase Electronic Phantom Load

Application

On site testing of all types of electricity meters using the voltages of the network;

Analysis of the voltage and current loads;

Advantages

Compact, light weight, user friendly menu;

Full control from the keyboard and LCD display on the panel, or from a PC via wireless, long distance, Bluetooth interface;

Automatic meter error measurements, using predefined Test Plans;

On-board Consult in Analysis Mode;

Powerful software with MySQL database;

Option – Measurement of Current Transformers and the System Error (meter + CT) with optional current probes;

Option – Control via RS232 of EE-300V-3 – a Three-Phase Voltage Generator, creating an independent Test System for laboratory or on-site measurement of electricity meters

Technical data

General

Power Supply: 85...265VAC, 47...63Hz (Overvoltage protection up to 500VAC)
Power Consumption: max 320VA
Dimensions WxDxH: 465x355x175mm
Weight: 13.6kg
Housing: Rugged plastic case
Operation temperature range: -10°C...+45°C
Relative humidity: <85% non condensing
Safety Tests: IEC61010-1-2002
Degree of protection: IP-20 (opened), IP-65 (closed)

Three-Phase Current Load

Current Range (per phase): 1mA...100A
Output Power (per phase): 60VA
Accuracy: ±0.2 %
Harmonic Distortion: <0.8 %
Phase Shift (for each phase): 0.0°...359.9°

Three-Phase Standard

Measuring Quantities: Volts, Amps, Phase Shift, P, Q, S, Wh, VARh, VAh, PF, Frequency
Measuring Modes: 1 Phase 2 Wire, 3 Phase 3 Wire Delta, 3 Phase 4 Wire Wye
Pulse Inputs: Scanning Head, Manual Button, KYZ,
High Frequency Input (up to 200kHZ)
Pulse output: 100 000 imp/kWh
Flash Memory: 200 tests reviewable on the LCD display

Standard Measurement Accuracy

Quantity Range Error
Voltage 30...500VAC ≤ ±0.05 %
5...30VAC ≤ ±0.05 %*
Current (direct) 50mA...100A ≤ ±0.1 %
1mA...50mA ≤ ±0.1 %*
Current Clamps 100A 100mA...100A ≤ ±0.2 %
Current Clamps 1000A 20A...2000A ≤ ±0.2 %
Active/Reactive energy (direct) 50mA...100A ≤ ±0.1 %
Apparent energy (direct) 50mA...100A ≤ ±0.1 %
Energy (CC 100A) 200mA...100A ≤ ±0.2/PF %
Energy (CC 1000A) 20A...1000A ≤ ±0.2/PF %
Phase shift 0.00°...359.9° 0.1°
Power Factor -1.000...+1.000 0.001
Frequency 40Hz...65Hz 0.01Hz
Temperature coefficient 0°...40° ≤ 30ppm/K

* Error related to the maximum value of the range