EE-100A-3S Three-Phase Current Load with Energy Standard Class 0.1 + EE-300V-3 Three-Phase Voltage Generator

Three-Phase Electronic Phantom Load

Application

The Test System consists of two units:

The Test System is used for testing electricity meters on-site or in a laboratory. It can generate three-phase voltages and currents, each phase being individually adjustable as amplitude and phase shift versus Va.

All types of electricity meters can be tested with the system – for active, reactive or apparent energy; 2-wire, 3-wire or 4-wire connection.

Advantages

Versatile structure with three main modes of operation:

The phase shifts and amplitudes of all voltages and currents are fully adjustable, which helps to simulate wrong connections or tamper;

Full control from the keyboard and the display on the panel, or from a PC via wireless long distance Bluetooth interface;

Electricity Error Test, Register Test, Starting Current, No Load Test;

Automatic meter error measurements using predefined Test Plans;

Powerful Software with MySQL Database

Technical Data

General

EE-100A-3S EE-300V-3
Power Supply: 85...265VAC, 47...63Hz 85...265VAC, 47...63Hz
(Overvoltage protection up to 500VAC)
Power Consumption: max 320VA max 130VA
Dimensions WxDxH: 465x355x175mm 465x355x175mm
Weight: 13kg 10,5kg
Housing: Rugged plastic case Rugged plastic case
Operation temperature range: -10°C...+45°C -10°C...+45°C
Relative humidity: <85% non condensing <85% non condensing
Safety Tests: IEC61010-1-2002 IEC61010-1-2002
Degree of protection: IP-20 (opened), IP-65 (closed) IP-20 (opened), IP-65 (closed)

Three-Phase Load

EE-100A-3S EE-300V-3
Range (per phase): 1mA...100A (step 1mA) 5V...300V (step 0,1V)
Output Power (per phase): 60VA 30VA
Accuracy: ±0.2 % ±0.2 %
Harmonic Distortion: <0.8 % <0.8 %
Phase Shift (for each phase): 0.0°...359.9° (step 0,1°) 0.0°...359.9° (step 0,1°)
(VA is always 0,0°)
Frequency (for all phases): 45...65Hz (step 0.1Hz) 45...65Hz (step 0.1Hz)

Three-Phase Standard

Measuring Quantities: Volts, Amps, Phase Shift, P, Q, S, Wh, VARh, VAh, PF, Frequency
Measuring Modes: 1 Phase 2 Wire, 3 Phase 3 Wire Delta, 3 Phase 4 Wire Wye
Pulse Inputs: Scanning Head, Manual Button, KYZ,
High Frequency Input (up to 200kHZ)
Pulse output: 100 000 imp/kWh
Flash Memory: 200 tests reviewable on the LCD display

Standard Measurement Accuracy

Quantity Range Error
Voltage 30...500VAC ≤ ±0.05 %
5...30VAC ≤ ±0.05 %*
Current (direct) 50mA...100A ≤ ±0.1 %
1mA...50mA ≤ ±0.1 %*
Active/Reactive energy (direct) 50mA...100A ≤ ±0.1/PF %
Apparent energy (direct) 50mA...100A ≤ ±0.1 %
Phase shift 0.00°...359.9° 0.1°
Power Factor -1.000...+1.000 0.001
Frequency 40Hz...65Hz 0.01Hz
Temperature coefficient 0°...40° ≤ 30ppm/K

* Error related to the maximum value of the range